Adverse reaction to metal debris following hip resurfacing
The Influence of Component Type, Orientation and Volumetric Wear
D. J. Langton, T. J. Joyce, S. S. Jameson, J. Lord, M. Van Orsouw, J. P. Holland, A. V. F. Nargol, K. A. De Smet
J Bone Joint Surg [Br] 2011;93-B:164-71.
We sought to establish the incidence of joint failure secondary to adverse reaction to metal debris (ARMD) following metal-on-metal hip resurfacing in a large, three surgeon, multicentre study involving 4226 hips with a follow-up of 10 to 142 months. Three implants were studied: the Articular Surface Replacement; the Birmingham Hip Resurfacing; and the Conserve Plus. Retrieved implants underwent analysis using a co-ordinate measuring machine to determine volumetric wear. There were 58 failures associated with ARMD. The median chromium and cobalt concentrations in the failed group were significantly higher than in the control group (p < 0.001). Survival analysis showed a failure rate in the patients with Articular Surface Replacement of 9.8% at five years, compared with < 1% at five years for the Conserve Plus and 1.5% at ten years for the Birmingham Hip Resurfacing. Two ARMD patients had relatively low wear of the retrieved components. Increased wear from the metal-on-metal bearing surface was associated with an increased rate of failure secondary to ARMD. However, the extent of tissue destruction at revision surgery did not appear to be dose-related to the volumetric wear.
- Predicting Dissatisfaction After THA
- Sex as a Patient Selection Criterion for Metal-on-Metal HRA
- Whole Blood Metal Ion Concentrations
- Magnetic Resonance Imaging Findings in Painful Metal-On-Metal Hips
- Adverse reaction to metal debris following hip resurfacing
- Ceramic-on-Ceramic Hip Outcome at a 5- to 10-Year Interval
- Wear Comparison
- The Incidence of Acetabular Osteolysis in Young Patients
- A Monoblock Porous Tantalum Acetabular Cup Has No Osteolysis on CT at 10 Years
- Large Acetabular Defects Can be Managed with Cementless Revision Components
- A Comparison of Two Implant Systems in Restoration of Hip Geometry in Arthroplasty
- Total Hip Arthroplasty Modular Neck Failure
- Revision of Failed THA Acetabular Cups to Porous Tantalum Components
- Metal-on-Metal Hip Arthroplasty in Patients Thirty Years of Age or Younger
- Metal-on-Metal Total Hip Arthroplasty
- Optimal acetabular orientation for hip resurfacing
- Histological Features of Pseudotumor-like Tissues From Metal-on-Metal Hips
- The Prevalence of Groin Pain After Metal-on-Metal THA and THR
- Risk of Complication and Revision Total Hip Arthroplasty