Polyethylene Subluxation: A Radiographic Sign of Locking Mechanism Failure After Modular TKA
Matthew S. Hepinstall, MD, and José A. Rodriguez, MD
The Journal of Arthroplasty, volume 26, issue 1
Abstract: Tibial insert locking mechanisms are intended to limit interface motion and “backside” wear in modular total knee arthroplasty (TKA). Nevertheless, anterior polyethylene subluxation is occasionally apparent on lateral radiographs after TKA, suggesting locking mechanism failure. We retrospectively identified 10 modular posterior-stabilized implants of a single design that were found to have failure of the locking mechanism at the time of revision surgery for osteolysis.
Operative reports were reviewed for the presence of backside wear, and preoperative radiographs were inspected for polyethylene subluxation. All 10 implants demonstrated significant backside wear. Nine had anterior polyethylene subluxation evident on preoperative radiographs. Anterior polyethylene subluxation on the lateral radiograph is a subtle sign of failure of the locking mechanism in this modular posterior-stabilized TKA. Keywords: total knee replacement, modularity, locking mechanism, backside wear, radiographic, evaluation.
- Pain and Function in Patients After Primary Unicompartmental and Total Knee Arthroplasty
- Complications After Minimally Invasive TKA as Compared With Traditional Incision Techniques
- Polyethylene Subluxation: A Radiographic Sign of Locking Mechanism Failure After Modular TKA
- Measuring Patellofemoral Forces and Pressures in a Simulated Operating Room Environment
- Skyline patellofemoral radiographs can only exclude late stage degenerative changes
- The John Insall Award
- HTA REPORT
- Wear and Lysis is the Problem in Modular TKA in the Young OA Patient at 10 Years
- TKA following medial opening wedge tibial osteotomy