In Vivo Oxidation in Remelted Highly Cross-Linked Retrievals
B.H. Currier, MChE, D.W. Van Citters, PhD, J.H. Currier, MS, and J.P. Collier, DE
J Bone Joint Surg Am. 2010;92:2409-18 d doi:10.2106/JBJS.I.01006
Background: Elimination of free radicals to prevent oxidation has played a major role in the development and product differentiation of the latest generation of highly cross-linked ultra-high molecular weight polyethylene bearing materials. In the current study, we (1) examined oxidation in a series of retrieved remelted highly cross-linked ultra-high molecular weight polyethylene bearings from a number of device manufacturers and (2) compared the retrieval results with findings for shelf-stored control specimens. The hypothesis was that radiation-cross-linked remelted ultra-high molecular weight polyethylene would maintain oxidative stability in vivo comparable with the stability during shelf storage and in published laboratory aging tests.
Methods: Fifty remelted highly cross-linked ultra-high molecular weight polyethylene acetabular liners and nineteen remelted highly cross-linked ultra-high molecular weight polyethylene tibial inserts were received after retrieval from twenty-one surgeons from across the U.S. Thirty-two of the retrievals had been in vivo for two years or more. Each was measured for oxidation with use of Fourier transform infrared spectroscopy. A control series of remelted highly crosslinked ultra-high molecular weight polyethylene acetabular liners from three manufacturers was analyzed with electron paramagnetic resonance spectroscopy to measure free radical content and with Fourier transform infrared spectroscopy to measure oxidation initially and after eight to nine years of shelf storage in air.
Results: The never-implanted, shelf-aged controls had no measurable free-radical content initially or after eight to nine years of shelf storage. The never-implanted controls showed no increase in oxidation during shelf storage. Oxidation measurements showed measurable oxidation in 22% of the retrieved remelted highly cross-linked liners and inserts after an average of two years in vivo.
Conclusions: Because never-implanted remelted highly cross-linked ultra-highmolecular weight polyethylenematerials had no measurable free-radical concentration and no increase in oxidation during shelf storage, thesematerials were expected to be oxidation-resistant in vivo. However, some remelted highly cross-linked ultra-highmolecular weight polyethylene retrievals showed measurable oxidation after an average of more than two years in vivo. This apparent departure from widely expected behavior requires continued study of the process of in vivo oxidation of ultra-high molecular weight polyethylene materials.
Clinical Relevance: Remelted highly cross-linked ultra-high molecular weight polyethylene acetabular and tibial retrievals showed unexpected oxidation.
- Rotational References for TKA Tibial Components Change with Level of Resection
- Femoral Notch Stenosis Caused by Soft Tissue Impingement in Semi or Open-Box Posterior-Stabilized TKA
- Do Porous Tantalum Implants Help Preserve Bone?
- Posterior cruciate ligament recruitment affects antero-posterior translation during flexion gap distraction in total knee replacement
- Long-Term Outcomes of Patellofemoral Arthroplasty
- Younger age increases the risk of early prosthesis failure following primary total knee replacement for osteoarthritis
- Modern Treatment of Infected TKA With a 2-Stage Reimplantation Protocol
- Limitations of the Knee Society Score in Evaluating Outcomes Following Revision TKA
- In Vivo Oxidation in Remelted Highly Cross-Linked Retrievals