Histological Features of Pseudotumor-like Tissues From Metal-on-Metal Hips
Pat Campbell PhD, Edward Ebramzadeh PhD, Scott Nelson MD, Karren Takamura BA, Koen De Smet MD, Harlan C. Amstutz MD
Clinical Orthopaedics and Related Research1 Volume 468, Number 9, September 2010
Background Pseudotumor-like periprosthetic tissue reactions around metal-on-metal (M-M) hip replacements can cause pain and lead to revision surgery. The cause of these reactions is not well understood but could be due to excessive wear, or metal hypersensitivity or an as-yet unknown cause. The tissue features may help distinguish reactions to high wear from those with suspected metal hypersensitivity.
Questions/purposes We therefore examined the synovial lining integrity, inflammatory cell infiltrates, tissue organization, necrosis and metal wear particles of pseudotumorlike tissues from M-M hips revised for suspected high wear related and suspected metal hypersensitivity causes.
Methods Tissue samples from 32 revised hip replacements with pseudotumor-like reactions were studied. A 10- point histological score was used to rank the degree of aseptic lymphocytic vasculitis-associated lesions (ALVAL) by examination of synovial lining integrity, inflammatory cell infiltrates, and tissue organization. Lymphocytes, macrophages, plasma cells, giant cells, necrosis and metal wear particles were semiquantitatively rated. Implant wear was measured with a coordinate measuring machine. The cases were divided into those suspected of having high wear and those suspected of having metal hypersensitivity based on clinical, radiographic and retrieval findings. The Mann-Whitney test was used to compare the histological features in these two groups.
Results The tissues from patients revised for suspected high wear had a lower ALVAL score, fewer lymphocytes, but more macrophages and metal particles than those tissues from hips revised for pain and suspected metal hypersensitivity. The highest ALVAL scores occurred in patients who were revised for pain and suspected metal hypersensitivity. Component wear was lower in that group.
Conclusions Pseudotumor-like reactions can be caused by high wear, but may also occur around implants with low wear, likely because of a metal hypersensitivity reaction. Histologic features including synovial integrity, inflammatory cell infiltrates, tissue organization, and metal particles may help differentiate these causes. Clinical Relevance Painful hips with periprosthetic masses may be caused by high wear, but if this can be ruled out, metal hypersensitivity should be considered.
- Predicting Dissatisfaction After THA
- Sex as a Patient Selection Criterion for Metal-on-Metal HRA
- Whole Blood Metal Ion Concentrations
- Magnetic Resonance Imaging Findings in Painful Metal-On-Metal Hips
- Adverse reaction to metal debris following hip resurfacing
- Ceramic-on-Ceramic Hip Outcome at a 5- to 10-Year Interval
- Wear Comparison
- The Incidence of Acetabular Osteolysis in Young Patients
- A Monoblock Porous Tantalum Acetabular Cup Has No Osteolysis on CT at 10 Years
- Large Acetabular Defects Can be Managed with Cementless Revision Components
- A Comparison of Two Implant Systems in Restoration of Hip Geometry in Arthroplasty
- Total Hip Arthroplasty Modular Neck Failure
- Revision of Failed THA Acetabular Cups to Porous Tantalum Components
- Metal-on-Metal Hip Arthroplasty in Patients Thirty Years of Age or Younger
- Metal-on-Metal Total Hip Arthroplasty
- Optimal acetabular orientation for hip resurfacing
- Histological Features of Pseudotumor-like Tissues From Metal-on-Metal Hips
- The Prevalence of Groin Pain After Metal-on-Metal THA and THR
- Risk of Complication and Revision Total Hip Arthroplasty