In Vivo Wear Rate of the Birmingham Hip Resurfacing Arthroplasty
A Review of 10 Retrieved Components
Wolf-Christoph Witzleb, MD, Uwe Hanisch, MD, Joerg Ziegler, MD, and Klaus-Peter Guenther, MD
The Journal of Arthroplasty Vol. 24 No. 6 2009
Abstract The wear of 8 femoral and 2 acetabular components of the Birmingham Hip Resurfacing (Midland Medical Technologies Ltd, Birmingham, UK) was measured using a coordinate measuring machine (CMM5; SIP, Geneva, Switzerland). After a median in situ time of 13 months, measurements of the Birmingham Hip Resurfacing femoral heads revealed a median volumetric wear of 2.9 mm3. In one case with a cup abduction angle of 70°, a significantly higher wear of 17.8 mm3 was found. The 2 acetabular components showed volumetric wear of 5.4 and 27.6 mm3 after 14 and 15 months in vivo. The wear rates of the investigated femoral components were somewhat, but not significantly, higher than the previously investigated 28-mm Metasul heads (median, 0.8 mm3; Zimmer GmbH, Winterthur, Switzerland).
- Results With the Cementless Spotorno Stem in Total Hip Arthroplasty
- Third-Generation Ceramic-on-Ceramic Bearing Surfaces in Revision Total Hip Arthroplasty
- Corrosion of a Hip Stem With a Modular Neck Taper Junction
- Elevated Serum Concentrations of Cobalt, Chromium, Nickel, and Manganese After Metal-On-Metal Alloarthroplasty of the Hip
- Long-term Results for the Uncemented Zweymuller/Alloclassic Hip Endoprosthesis
- In Vivo Wear Rate of the Birmingham Hip Resurfacing Arthroplasty
- The management of severe acetabular bone defects in revision hip arthroplasty using modular porous metal components
- Cementation of a Polyethylene Liner Into a Metal Acetabular Shell
- Blood metal ion concentrations after hip resurfacing arthroplasty
- Migration of the Trabecular Metal Monoblock Acetabular Cup System
- Risk factors for inflammatory pseudotumour formation following hip resurfacing
- Primary Total Hip Arthroplasty With Duraloc Cup in Patients Younger Than 50 Years
- Ten-year follow-up of the non-porous Allofit cementless acetabular component
- Ceramic-on-ceramic bearing surface and risk of revision due to dislocation after primary total hip replacement